Loading…
Name ID Label Description Type Access Class
var_TB82_var_ref_temp 4024 Variable ref temp Variable Reference Temperature provides the ability to use use any temp as the reference temp used to determine the temp-compensated conductivity value (i.e., Primary value). R/W Float 8
var_TB82_user_defined_tc_t1 4048 Temp 1 Temp 1 of user defined temperature coefficient. First end point of the curve. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t2 4049 Temp 2 Temp 2 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t3 404A Temp 3 Temp 3 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t4 404B Temp 4 Temp 4 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t5 404C Temp 5 Temp 5 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t6 404D Temp 6 Temp 6 of user defined temperature coefficient. Last end point of the curve. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k1 404E K1/Kref Ratio of uncompensated conductivity at Temp 1 to conductivity at reference temperature (K1/Kref). First endpoint of the curve. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k2 404F K2/Kref Ratio of uncompensated conductivity at Temp 2 to conductivity at reference temperature (K2/Kref). Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k3 4050 K3/Kref Ratio of uncompensated conductivity at Temp 3 to conductivity at reference temperature (K3/Kref). Valid only when Auto-TC option is user-defined R/W Float 8
Name ID Label Description Type Access Class