Name ID Label Description Type Access Class
var_TB82_func_gen_x5_val 4023 Spcl xfer curve x5 Transfer Function Special Curve x5 value (input value point 5 entered as a percentage of Primary value range). Valid only when PV xfer fnctn is spcl curve R/W Float 8
var_TB82_func_gen_y1_val 401C Spcl xfer curve y1 Transfer Function Special Curve y1 value (Analog output value point 1 entered as a percentage of Analog Output). Valid only when PV xfer fnctn is spcl curve R/W Float 8
var_TB82_func_gen_y2_val 401E Spcl xfer curve y2 Transfer Function Special Curve y2 value (Analog output value point 2 entered as a percentage of Analog Output). Valid only when PV xfer fnctn is spcl curve R/W Float 8
var_TB82_func_gen_y3_val 4020 Spcl xfer curve y3 Transfer Function Special Curve y3 value (Analog output value point 3 entered as a percentage of Analog Output). Valid only when PV xfer fnctn is spcl curve R/W Float 8
var_TB82_func_gen_y4_val 4022 Spcl xfer curve y4 Transfer Function Special Curve y4 value (Analog output value point 4 entered as a percentage of Analog Output). Valid only when PV xfer fnctn is spcl curve R/W Float 8
var_TB82_func_gen_y5_val 4024 Spcl xfer curve y5 Transfer Function Special Curve y5 value (Analog output value point 5 entered as a percentage of Analog Output). Valid only when PV xfer fnctn is spcl curve R/W Float 8
var_TB82_lcd_secondary 402D LCD secondry disp The parameter sets the HMI's local LCD secondary display area to one of several options. R/W Enum 8
var_TB82_security_lock 402E Security Lock When ENABLED, the Security Password will be active for Security mode on local HMI. If Security Lock is disabled, then the Security mode of operation can be accessed from the local HMI without supplying the password. R/W Enum 8
var_TB82_configuration_lock 402F Configuration Lock When ENABLED, the Security Password will be active for Config mode on local HMI. Config parameters will be write protected on the HART interface. When the Configuration-Lock is DISABLED, the Config mode of operation can be accessed from the local HMI without supplying the password R/W Enum 8
var_TB82_calibration_lock 4030 Calibration Lock When ENABLED, the Security Password will be active for Calibrate mode on local HMI. Calib parameters will be write protected on the HART interface. When the Calibration Lock is DISABLED, the Calibrate mode of operation can be accessed from the local HMI without supplying the password R/W Enum 8
Name ID Label Description Type Access Class