Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_TB82_temp_comp_type | 4017 | Temp comp type | Determines the type of Temperature Compensation algorithm applied to the Primary Value | R/W | Enum | 8 |
var_TB82_manual_temp_setpoint_unit | 4044 | Manual temp unit | R/W | Enum | 8 | |
var_TB82_manual_temp_setpoint | 401A | Manual Temp. | Manual Temperature Compensation Value - is used in Nernstian temp. comp. when the Temp Comp selection is Manual | R/W | Float | 8 |
var_TB82_ph_10c | 4018 | pH solution coeff | The solution coefficient is in terms of pH units per 10 deg-C for the pH analyzer state and has a range of +/- 9.99 pH units per 10 deg-C. The solution coeff is used only when Auto-Solution temp comp is selected. | R/W | Float | 8 |
var_TB82_mv_10c | 4019 | mV solution coeff | The solution coefficient is in terms of millivolts per 10 deg-C for the ORP, pION and ion-conc states and has a range of +/-19.99 mV per 10 deg-C. The solution coeff is used only when Auto-Solution temp comp is selected. | R/W | Float | 8 |
info_text1 | 4062 | Enter all the curve points and then send the curve to the device | Information indicating not to transmit individual curve points one at a time to the device | R | ASCII | 4096 |
var_TB82_func_gen_x1_val | 401B | Spcl xfer curve x1 | Transfer Function Special Curve x1 value (input value point 1 entered as a percentage of Primary value range). Valid only when PV xfer fnctn is spcl curve | R/W | Float | 8 |
var_TB82_func_gen_x2_val | 401D | Spcl xfer curve x2 | Transfer Function Special Curve x2 value (input value point 2 entered as a percentage of Primary value range). Valid only when PV xfer fnctn is spcl curve | R/W | Float | 8 |
var_TB82_func_gen_x3_val | 401F | Spcl xfer curve x3 | Transfer Function Special Curve x3 value (input value point 3 entered as a percentage of Primary value range). Valid only when PV xfer fnctn is spcl curve | R/W | Float | 8 |
var_TB82_func_gen_x4_val | 4021 | Spcl xfer curve x4 | Transfer Function Special Curve x4 value (input value point 4 entered as a percentage of Primary value range). Valid only when PV xfer fnctn is spcl curve | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |