Name ID Label Description Type Access Class
pv_units 4000 PV eng. unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R Enum 8
pv_value 4001 PV Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R Float 10
sensor_upper_range_limit 4002 PV USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
sensor_lower_range_limit 4003 PV LSL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
sensor_minimum_span 4004 PV Min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
pv_damping_value 4005 PV Damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 8
analog_value 4006 AO Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes R/W Float 66
fail_mode 4007 AO Alrm typ Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. R Enum 64
percent_range 4008 PV % rnge Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34
transfer_function 1F84 Xfer fnctn Transfer Function- Defines the transformation function that will be applied from the Field Device Variable to the Analog Output, and Percent Range. The Transfer Function does not affect the Digital Value representation. R Enum 32
pv_upper_range_value 4009 PV URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
pv_lower_range_value 400A PV LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
sensor_serial_number 400B PV Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
temp_unit 400C Temp Unit R Enum 8
temp_value 400D Temp Temperature- In Auto temp comp mode, it is the Temp read with the temp sensor. In Manual temp comp mode, it is the user entered temp value R Float 10
ref_impedance_unit 400E Ref Impedance unit R Enum 8
ref_impedance_value 400F Ref. Impedance Reference Impedance R Float 10
input_mv_unit 4010 Input unit R Enum 8
input_mv_value 4011 Input Input mV - Millivolt input value. R Float 10
PV_efficiency 4024 PV efficiency PV sensor efficiency (slope) value in percent R/W Float 8
PV_offset 4025 PV offset PV sensor offset value. R/W Float 8
temp_slope 4026 Temperature sensor Slope Temperature sensor Slope R/W Float 8
temp_offset 4027 Temperature sensor offset Temperature sensor offset value. R/W Float 8
temp_cal_2pt 401C 2 Point Cal Temp. value Temperature value for 2 point PV calib W Float 8
PV_cal_low_value 4018 PV cal low value PV low value for 2 point calibration. W Float 8