Name ID Label Description Type Access Class
upperRange_value 00EA PV URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
lowerRange_value 00EB PV LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
pv_Value 4036 PV R Float 10
pv_Units 4037 PV Snsr unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 8
pv_USL 400B USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
pv_LSL 400A LSL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
pv_Min_Span 400C Minimum Span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
pv_Damp 4038 Damping Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 8
pv_SSN 4007 Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
analog_value 400D AO Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes R/W Float 66
fail_mode 400E AO Alrm typ Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. R Enum 64
transfer_function 1F84 Xfer fnctn Transfer Function- Defines the transformation function that will be applied from the Field Device Variable to the Analog Output, and Percent Range. The Transfer Function does not affect the Digital Value representation. R Enum 32
percent_range 4039 PV % rnge Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34
sv_Value 403A Temp (SV) Secondary variable temperature R Float 10
sv_Units 403B Temp (SV) unit Secondary variable temperature engineering units R/W Enum 8
tv_Value 403C Cond/Conc (TV) Device Tertiary Variable (TV). TV = Concentration value when PV type = Conductivity. TV = Conductivity value when PV type = Concentration. R Float 10
tv_Units 403D Cond/Conc (TV) unit Tertiary variable conductivity or concentration engineering units R/W Enum 8
qv_Value 403E Uncomp Cond (QV) Quaternary variable temp-uncompensated conductivity R Float 10
qv_Units 403F Uncomp Cond (QV) unit Quaternary variable uncompensated conductivity engineering units R/W Enum 8
var_TB82_primary_value_type 400F PV type Primary Value (PV) type selection. When PV type = Conductivity, PV value = Conductivity and TV value = Concentration. When PV type = Concentration, PV value = Concentration and TV value = Conductivity. R/W Enum 8
var_TB82_sensor_group 4010 Snsr group Three sensor groups are applied and defined by the sensor cell constant: A, B and C. R/W Enum 8
var_TB82_sensor_diagnostics 4021 Snsr diagnostics Sensor Diagnostics selection determines whether certain sensor diagnostics are enabled or disabled R/W Enum 8
var_TB82_solution 4011 Solution type The solution type defines the solute algorithm used for concentration configurations. Both predefined and user-defined concentration configurations are available. The available solution types are dependent upon the current SENSOR_GROUP. Sensor Group A: 0 to 15% NaOH, 0 to 20% NaCl, 0 to 18% HCl, 0 to 20% H2SO4, User Defined. Sensor Group B: User Defined. Sensor Group C: User Defined. R/W Enum 8
var_TB82_usrdef_secondary 4012 Concen curve name This is a six character text name for the concentration curve. It could be modified in user-defined concentration configuration R/W ASCII 2048
var_TB82_usrdef_unit 4013 User def conc unit Engineering units for user defined concentration curve This will be used as the concentration value units when a user defined concentration curve is selected. R/W Enum 8
Name ID Label Description Type Access Class