Loading…
名称 ID 标签 描述 类型 访问 分类
var_TB82_user_defined_conc_x5 401D x5 Cond x5 Conductivity point of the user defined concentration curve. 读/写 浮点 8
var_TB82_user_defined_conc_x6 401F x6 Cond x6 Conductivity point of the user defined concentration curve. 读/写 浮点 8
var_TB82_user_defined_conc_y1 4016 y1 Concen y1 Concentration point corresponding to x1 Conductivity in the user defined concentration curve 读/写 浮点 8
var_TB82_user_defined_conc_y2 4018 y2 Concen y2 Concentration point corresponding to x1 Conductivity in the user defined concentration curve 读/写 浮点 8
var_TB82_user_defined_conc_y3 401A y3 Concen y3 Concentration point corresponding to x1 Conductivity in the user defined concentration curve 读/写 浮点 8
var_TB82_user_defined_conc_y4 401C y4 Concen y4 Concentration point corresponding to x1 Conductivity in the user defined concentration curve 读/写 浮点 8
var_TB82_user_defined_conc_y5 401E y5 Concen y5 Concentration point corresponding to x1 Conductivity in the user defined concentration curve 读/写 浮点 8
var_TB82_user_defined_conc_y6 4020 y6 Concen y6 Concentration point corresponding to x1 Conductivity in the user defined concentration curve 读/写 浮点 8
var_TB82_lcd_secondary 402D LCD secondry disp The parameter sets the HMI's local LCD secondary display area to one of several options. 读/写 枚举 256
var_TB82_security_lock 402E Security Lock When ENABLED, the Security Password will be active for Security mode on local HMI. If Security Lock is disabled, then the Security mode of operation can be accessed from the local HMI without supplying the password. 读/写 枚举 2048
var_TB82_configuration_lock 402F Configuration Lock When ENABLED, the Security Password will be active for Config mode on local HMI. Config parameters will be write protected on the HART interface. When the Configuration-Lock is DISABLED, the Config mode of operation can be accessed from the local HMI without supplying the password 读/写 枚举 2048
var_TB82_calibration_lock 4030 Calibration Lock When ENABLED, the Security Password will be active for Calibrate mode on local HMI. Calib parameters will be write protected on the HART interface. When the Calibration Lock is DISABLED, the Calibrate mode of operation can be accessed from the local HMI without supplying the password 读/写 枚举 2048
var_TB82_output_lock 4031 Output Lock When ENABLED, the Security Password will be active for Output mode on local HMI. Output related parameters will be write protected on the HART interface. When the Output Lock is DISABLED, the Output mode of operation can be accessed from the local HMI without supplying the password 读/写 枚举 2048
var_TB82_password_digit_1 4081 Secur. Passwrd 1 Three password digits form the Security Password. Each digit can have a value from 0 to 9. The 3-digit Password range is 000 to 999. When Secur, Config, Output or Calib Locks are ENABLED, this Security Password will be required at the local HMI to access the corresponding modes of operation. 读/写 无符号 2048
var_TB82_password_digit_2 4082 Secur. Passwrd 2 Three password digits form the Security Password. Each digit can have a value from 0 to 9. The 3-digit Password range is 000 to 999. When Secur, Config, Output or Calib Locks are ENABLED, this Security Password will be required at the local HMI to access the corresponding modes of operation. 读/写 无符号 2048
var_TB82_password_digit_3 4083 Secur. Passwrd 3 Three password digits form the Security Password. Each digit can have a value from 0 to 9. The 3-digit Password range is 000 to 999. When Secur, Config, Output or Calib Locks are ENABLED, this Security Password will be required at the local HMI to access the corresponding modes of operation. 读/写 无符号 2048
var_TB82_pv_slope_units 4062 PV slope units 读/写 枚举 8
var_TB82_pv_slope 4029 PV Slope PV sensor efficiency (slope) value in percent 读/写 浮点 8
var_TB82_pv_offset_units 4045 PV snsr offset unit PV sensor offset unit 读/写 枚举 8
var_TB82_pv_offset 402A PV offset PV sensor offset value. 读/写 浮点 8
var_TB82_temp_slope_units 4065 Temp slope units 读/写 枚举 8
var_TB82_temp_slope 402B Temp snsr Slope Temperature sensor Slope 读/写 浮点 8
var_TB82_temp_offset_units 4047 Temp snsr offset unit Temperature sensor offset unit 读/写 枚举 8
var_TB82_temp_offset 402C Temp snsr offset Temp snsr offset 读/写 浮点 8
var_TB82_cal_status 4088 Cal Status 枚举 8
var_TB82_temp_cal_unit 4089 Temp cal unit 读/写 枚举 8
var_TB82_temp_cal_value 408A Temp cal value 读/写 浮点 8
var_TB82_one_point_PV_value 4092 PV value for PV cal 读/写 浮点 8
var_TB82_pv_cal_unit 4093 读/写 枚举 8
unique_device_type_code 07ED 读/写 枚举 4096
rosemount_model_code 07F5 读/写 枚举 4096
rosemount_analytical_model_code 07F6 读/写 枚举 4096
kayray_model_code 07F7 读/写 枚举 4096
micro_motion_model_code 07F8 读/写 枚举 4096
units_code 07E9 读/写 枚举 4096
transfer_function_code 07EA 读/写 枚举 4096
material_code 07EC 读/写 枚举 4096
alarm_selection_code 07EB 读/写 枚举 4096
write_protect_code 07EE 读/写 枚举 4096
company_identification_code 07EF 读/写 枚举 4096
burst_mode_control_code 07F0 读/写 枚举 4096
physical_signaling_codes 07F1 读/写 枚举 4096
flag_assignment 07F2 读/写 枚举 4096
operating_mode_code 07F3 读/写 枚举 4096
analog_output_numbers_code 07F4 读/写 枚举 4096
response_code 0096 特定命令响应代码——现场设备响应返回的数据,表示主机请求命令发送成功。该信息可能与主机请求数据或命令本身有关。 读/写 枚举 129
comm_status 0098 通讯错误汇总——现场设备在尝试接收主机请求时检测到下列错误条件。因此命令未被接受,并且主机请求在现场设备上未引起变化。 读/写 位枚举 131
device_status 0097 现场设备状态——表示现场设备已检测到有关硬件/变量有效性/操作状态/内部过程的条件的数据。 读/写 位枚举 131
manufacturer_id 0099 厂商 厂商ID代码——指特殊厂商,一般为制造该现场设备的公司名。 枚举 128
device_type 009A 型号 型号——指现场设备类型,一般是单一厂商唯一的广告型号。 枚举 128
名称 ID 标签 描述 类型 访问 分类