Name ID Label Description Type Access Class
pv_units 4000 PV eng. unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 8
pv_value 4002 PV Conductivity - Temperature compensated process solution conductivity. R Float 10
sensor_upper_range_limit 4003 PV USL Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. R Float 8
sensor_lower_range_limit 4004 PV LSL Lower Sensor Limit Value- Defines the minimum usable value for the Lower Range Value. R Float 8
sensor_minimum_span 4005 PV Min span Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. R Float 8
pv_damping_value 4006 PV Damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R/W Float 8
analog_value 4007 AO Analog Output Value- Value that tracks the Digital Value representation, under normal operating modes R/W Float 66
fail_mode 4008 AO Alrm typ Analog Output Alarm Type- Defines how the Analog Output will respond when the Field Device detects that the Analog Output may not be tracking the associated Field Device Variable. NOTE- The Digital Value representation may not be determinate. R Enum 64
percent_range 4009 PV % rnge Percent of Range- Variable that tracks the Digital Value representation with respect to the range defined by the Lower Range Value and Upper Range Value, for normal operating modes. The units of this variable are always in percent. R Float 34
transfer_function 1F84 Xfer fnctn Transfer Function- Defines the transformation function that will be applied from the Field Device Variable to the Analog Output, and Percent Range. The Transfer Function does not affect the Digital Value representation. R Enum 32
pv_upper_range_value 400A PV URV Upper Range Value- Defines the operational endpoint from which the Analog Value, and the 100% point of the Percent Range are derived. In addition, the Upper Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
pv_lower_range_value 400B PV LRV Lower Range Value- Defines the operational endpoint from which the Analog Value, and the 0% point of the Percent Range are derived. In addition, the Lower Range Value defines an operational endpoint which the alarms associated with the Analog Value, and the alarms associated with the Digital Value representation are derived. R/W Float 32
sensor_serial_number 400C PV Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R/W Unsigned 2048
temp_unit 400D Temp Unit R Enum 8
temp_value 400E Temp Temperature- In Auto temp comp mode, it is the Temp read with the temp sensor. In Manual temp comp mode, it is the user entered temp value R Float 10
PV_efficiency 4018 PV slope PV sensor slope value R/W Float 8
PV_offset 4019 PV offset PV sensor offset value. R/W Float 8
temp_slope 401A Temperature sensor Slope Temperature sensor Slope R/W Float 8
temp_offset 401B Temperature sensor offset Temperature sensor offset value. R/W Float 8
one_point_PV_capture 4012 Cond cal value R Float 8
one_point_PV_value 4011 Cond cal value Conductivity calibration value W Float 8
temp_cal_value 4014 Temp. calibration value Temperature calibration value W Float 8
units_code 07E9 R/W Enum 4096
transfer_function_code 07EA R/W Enum 4096
alarm_selection_code 07EB R/W Enum 4096
Name ID Label Description Type Access Class