Name ID Label Description Type Access Class
uiEinheitQv 4010 Unit Qmax Selection of the units for the volume and mass flowrate measurements. The units are used for Range max., Qmax and the instantaneous display value when direct reading engineering units have been selected for the display. R/W Enum 2048
uiEinheitZv 4011 Unit totalizer Selections of the engineering totalizer unit for the volume and mass flowrate measurements. R/W Enum 2048
ucProgEinheitDichte 4036 Prog. Unit Configurable units for mass (with density) or volume flowrate (without density). R/W Enum 0
ucSchaltausgang 402E Contact output The switch mode for the contact output (terminals 41/42) can be selected as normally open or normally closed. R/W Enum 2048
ucSchalteingang 402F Contact input Selection of the function for the contact input terminals 81/82. R/W Enum 2048
ucIoutAlarm 404B Iout at Alarm Value of the current during an error condition. R/W Enum 2048
ucDlrEinAus 4038 Detector e.pipe On/Off selection for the operation Detector empty pipe. R/W Enum 2048
ucAlarmlRohr 4039 Alarm e.pipe When the meter tube is empty an indication is displayed and a signal is sent to the terminals when the alarm setting is on. No alarm contact when the Alarm setting is off. R/W Enum 2048
ucIoutlRohr 403A Iout at e.pipe Current output value Low or High during an empty pipe condition. R/W Enum 2048
ucSimulationEin 408D Simulation Mode Turn flowrate simulation on /off. R/W Enum 2048