Name ID Label Description Type Access Class
ucNetzfrequenz 40A4 Line frequency Selection of the line frequency of the flowmeter primary was calibrated. Selection 50/60 Hz only available for Older Models. R/W Enum 2048
ucStartUpKommuService 40A6 Startup Commu. With a TTL-Signal the communication type HART or ASCII can be selected of the service plug for the connection. R/W Enum 2048
ucTestImpulsausgangEin 40A8 Startup Commu. Function test for the pulse output. R/W Enum 2048
ucIoutArt 40AA Current output Selection of the current output range. For HART-Protocol always 4-20 mA. R/W Enum 2048
ucKommuServiceStecker 40AC Com. Type TTL The choice of the occupancy of the communication for the service plug can be returned. Communication HART- TTL or ASCII-TTL can be switched over to on display operate function. R/W Enum 2048
ucImpulsZustand 40AE Pulse (act/pas) Indication of the jumper position for pulse output active or passive. R/W Enum 2048
ucAktuellVariante 40B0 Actu. Variant The present factory-made for pulse output and communication. it can be identified by the order. R/W Enum 2048
ucSchalteingangTest 40B2 Contact input The status condition of the contact input can be viewed on or off. R/W Enum 2048
ucSchaltausgangTest 40B4 Contact output The test function contact ouput and with ENTER the function can be switched to on or off: R/W Enum 2048
ucDlrArt_LNS 403B EPD Mode Selection of the function standard or new adjustment. At standard is isn't required to carry out a adjustment on side. The converter use standard adjustment datas. If it comes to disturbances through the medium or an empty pipe isn't recognized by the converter, a new adjustment must be carried out on side, see adjustment empty pipe or full pipe. R/W Enum 2048