Name ID Label Description Type Access Class
pressure_value 138E 圧力 Field Device Variable Value- Digital value that can change without user intervention. Pressure- Measured proccess pressure in relation to a reference pressure. R Float 10
pressure_percent_range 138F 圧力 % レンジ Percent of Range- Variable which represents the digital value with respect to the range defined by the lower range value and upper range value. For normal operating modes. This variable is always in percent. Pressure- Measured process pressure in relation to a reference pressure. R Float 34
pressure_upper_range_limit 1393 圧力 USL Upper Sensor Limit Value- Defines the maximum value able to be used for the upper range value. Pressure- Measured proccess pressure in relation to a reference pressure. R/W Float 8
pressure_lower_range_limit 1394 圧力 LSL Lower Sensor Limit Value- Defines the lowest allowable value for the lower range value. Pressure- Pressure of the process with respect to a reference pressure. R/W Float 8
pressure_minimum_span 1390 圧力最小スパン Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. Pressure- Measured proccess pressure in relation to a reference pressure. R/W Float 8
pressure_output_transfer_function 1391 圧力伝送機能 伝送タイプ - フィールド機器のアナログ出力とパーセント レンジの伝送タイプを定義します。これはデジタル値表示には影響しません。 R/W Enum 32
pressure_upper_range_value 1396 圧力 URV Upper Range Value- Defines the endpoint for the analog value and the 100% point of the percent range. The upper range value also defines the operational endpoint from which the alarms for the analog output value as well as the associated digital value are taken. Pressure- Measured proccess pressure in relation to a reference pressure. R/W Float 34
pressure_lower_range_value 1397 圧力 LRV Lower Range Value- Defines the operational endpoint or the analog output signal and 0% of the percent range. In addition, the lower range value defines an operational endpoint from which the alarm values for the analog output as well as the associated digital value are derived. Pressure- Pressure of the process with respect to a reference pressure. R/W Float 34
pressure_damping_value 1398 圧力ダンピング Damping Value- Damping time which affects both the digital and analog output values. The variable is always in seconds. Pressure- Pressure of the process with respect to a reference pressure. R/W Float 10
flange_type 139A フランジ タイプ Flange Type- Part next to the sensor which physically connects the process to the sensor. Pressure- Pressure of the process measured with respect to a reference pressure. R/W Enum 2048
Name ID Label Description Type Access Class