Name ID Label Description Type Access Class
var_tempARStatus 4032 TC Recognition Status Temp sensor recognition status. Device scans for TC at power up, when TC type is switched from MANUAL to AUTOMATIC, and when TC recognition is run through the appropriate menu. R Enum 2
var_pureH2Ooption 402E Pure H2O option When Temp comp option is set to Pure H2O, temp comp is based on Pure H2O option selection of either a neutral salt, acid or base. R/W Enum 0
var_tempAutoTCOption 402B Auto TC Option Determines the algorithm that adjusts for the temperature effect of a solute. R/W Enum 8
var_tempAutoTCCoeff 402D TC Coeff Temp comp is based on percent change of conductivity from an adjustable reference temperature per deg C. The value is adjustable between 0 and 9.99 % per deg-C R/W Float 8
var_tempCompType 4029 Temp.Comp.Type Determines the type of Temperature Compensation algorithm applied to the Primary Value R/W Enum 8
var_manualTemp 402A Manual Temp Manual Temperature Compensation Value - is used in Nernstian temp. comp. when the Temp Comp selection is Manual R/W Float 8
var_refTemp 402F Reference Temp Variable reference temperature - this provides the ability to use any temp as the reference temp used to determine the temp-compensated PV R/W Float 8
var_TCUsrDefX1 40A9 Temp 1 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8
var_TCUsrDefX2 40AA Temp 2 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8
var_TCUsrDefX3 40AB Temp 3 When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. R/W Float 8