名称 ID 标签 描述 类型 访问 分类
var_concUnit 4023 Conc. Unit Concentration unit is % for pre-defined solute algorithms. unit can be selected between percent, ppm, µg/l, ppb, mg/l and none for user-defined solute algorithm. 读/写 枚举 8
var_hwWriteProtection 403B HW write protection Hardware write protection 枚举 2
var_swWriteProtection 403C SW write protection Software write protection 读/写 枚举 8
var_tempSensorType 4030 Temp. sensor type Temperature sensor type recognized 枚举 2
var_tempARStatus 4032 TC Recognition Status Temp sensor recognition status. Device scans for TC at power up, when TC type is switched from MANUAL to AUTOMATIC, and when TC recognition is run through the appropriate menu. 枚举 2
var_tempAutoTCOption 402C Auto TC Option Determines the algorithm that adjusts for the temperature effect of a solute. 读/写 枚举 8
var_tempAutoTCCoeff 402E TC Coeff Temp comp is based on percent change of conductivity from an adjustable reference temperature per deg C. The value is adjustable between 0 and 9.99 % per deg-C 读/写 浮点 8
var_tempCompType 402A Temp.Comp.Type Determines the type of Temperature Compensation algorithm applied to the Primary Value 读/写 枚举 8
var_manualTemp 402B Manual Temp Manual Temperature Compensation Value - is used in Nernstian temp. comp. when the Temp Comp selection is Manual 读/写 浮点 8
var_refTemp 402F Reference Temp Variable reference temperature - this provides the ability to use any temp as the reference temp used to determine the temp-compensated PV 读/写 浮点 8
名称 ID 标签 描述 类型 访问 分类