名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
var_analyzerType | 4015 | Analyzer type | PV measurement will be Conductivity or Concentration based on selection from Analyzer Type When PV is Concentration, Conductivity value is available as QV (Comp. Conductivity) | 读/写 | 枚举 | 8 |
var_sensorGroup | 4027 | Sensor Group | Sensor Group A or B based on sensor cell constant | 读/写 | 枚举 | 8 |
var_sensorDiag | 4028 | Sensor diag. | Sensor Diagnostics selection determines whether certain sensor diagnostics are enabled or disabled | 读/写 | 枚举 | 8 |
var_concSolution | 4022 | Conc. Solution | Defines the solute algorithm used for Concentration configuration Both predefined and user-defined concentration configurations are available. The available solution types are dependent upon the current Sensor Group. Sensor Group A: 0-15% NaOH, 0-20% NaCl, 0-18% HCl, 0-20% H2SO4, User Defined. Sensor Group B: User Defined. | 读/写 | 枚举 | 8 |
var_concName | 4024 | Conc. Curve Name | This is a six character text name for the concentration curve. It could be modified in user-defined concentration configuration. | 读/写 | ASCII | 8 |
var_concUnit | 4023 | Conc. Unit | Concentration unit is % for pre-defined solute algorithms. unit can be selected between percent, ppm, µg/l, ppb, mg/l and none for user-defined solute algorithm. | 读/写 | 枚举 | 8 |
var_hwWriteProtection | 403B | HW write protection | Hardware write protection | 读 | 枚举 | 2 |
var_swWriteProtection | 403C | SW write protection | Software write protection | 读/写 | 枚举 | 8 |
var_tempSensorType | 4030 | Temp. sensor type | Temperature sensor type recognized | 读 | 枚举 | 2 |
var_tempARStatus | 4032 | TC Recognition Status | Temp sensor recognition status. Device scans for TC at power up, when TC type is switched from MANUAL to AUTOMATIC, and when TC recognition is run through the appropriate menu. | 读 | 枚举 | 2 |
var_tempAutoTCOption | 402C | Auto TC Option | Determines the algorithm that adjusts for the temperature effect of a solute. | 读/写 | 枚举 | 8 |
var_tempAutoTCCoeff | 402E | TC Coeff | Temp comp is based on percent change of conductivity from an adjustable reference temperature per deg C. The value is adjustable between 0 and 9.99 % per deg-C | 读/写 | 浮点 | 8 |
var_tempCompType | 402A | Temp.Comp.Type | Determines the type of Temperature Compensation algorithm applied to the Primary Value | 读/写 | 枚举 | 8 |
var_manualTemp | 402B | Manual Temp | Manual Temperature Compensation Value - is used in Nernstian temp. comp. when the Temp Comp selection is Manual | 读/写 | 浮点 | 8 |
var_refTemp | 402F | Reference Temp | Variable reference temperature - this provides the ability to use any temp as the reference temp used to determine the temp-compensated PV | 读/写 | 浮点 | 8 |
var_TCUsrDefX1 | 40A9 | Temp 1 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefX2 | 40AA | Temp 2 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefX3 | 40AB | Temp 3 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefX4 | 40AC | Temp 4 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefX5 | 40AD | Temp 5 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefX6 | 40AE | Temp 6 | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefY1 | 40C5 | K1/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefY2 | 40C6 | K2/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefY3 | 40C7 | K3/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
var_TCUsrDefY4 | 40C8 | K4/Kref | When the user defined Auto temperature compensation is selected, the values in this curve are used for temperature compensation.The curve points Temp 1 to Temp 6 are the temperature values and the curve points K1/K25 to K6/K25 are the corresponding ratiosof uncompensated conductivity at Temp x to the conductivity at the reference temperature. | 读/写 | 浮点 | 8 |
名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |