名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
qv_Value | 4083 | Comp. Conductivity (QV) | Quaternary variable Comp. Conductivity | 读 | 浮点 | 10 |
qv_Units | 40A4 | Comp. Conductivity (QV) unit | Quaternary variable Comp. Conductivity engineering units | 读 | 枚举 | 8 |
hwRevisionText | 4011 | Hardware revision | Field device HW revision | 读 | ASCII | 2048 |
swRevisionText | 4012 | Software revision | Field device Firmware revision | 读 | ASCII | 2048 |
deviceSerialNumberText | 40D0 | Device Serial Number | Unique device serial No. | 读 | ASCII | 2048 |
var_analyzerType | 4015 | Analyzer type | PV measurement will be Conductivity or Concentration based on selection from Analyzer Type When PV is Concentration, Conductivity value is available as QV (Comp. Conductivity) | 读/写 | 枚举 | 8 |
var_sensorGroup | 4027 | Sensor Group | Sensor Group A or B based on sensor cell constant | 读/写 | 枚举 | 8 |
var_sensorDiag | 4028 | Sensor diag. | Sensor Diagnostics selection determines whether certain sensor diagnostics are enabled or disabled | 读/写 | 枚举 | 8 |
var_concSolution | 4022 | Conc. Solution | Defines the solute algorithm used for Concentration configuration Both predefined and user-defined concentration configurations are available. The available solution types are dependent upon the current Sensor Group. Sensor Group A: 0-15% NaOH, 0-20% NaCl, 0-18% HCl, 0-20% H2SO4, User Defined. Sensor Group B: User Defined. | 读/写 | 枚举 | 8 |
var_concName | 4024 | Conc. Curve Name | This is a six character text name for the concentration curve. It could be modified in user-defined concentration configuration. | 读/写 | ASCII | 8 |