Name ID Label Description Type Access Class
up_down_scale_mode 4001 Up-Down Scale mode Up-Down Scale mode - Defines the setting of the current output when fatal errors are detected. R/W Enum 2048
output_transfer_function 4002 Função Xfer Transfer Function - Defines the transformation function that will be applied between the Field Device Variable and the respective assigned Analog Output. The transfer function does not affect the Digital Value. R/W Enum 32
ameter_installation 4005 Tipo de metro Meter Installation - Defines whether there is a meter installed or not and, if integral, what information the meter will display. R/W Enum 256
push_buttons 4006 Push Buttons Push Buttons - Indicates whether the instrument is fitted with Zero and/or Span calibration buttons. R/W Enum 2048
aflange_type 4007 Tipo de flange Flange Type - Hardware, adjacent to the sensor, that physically connects the process to the sensor. R/W Enum 2048
aflange_material 4008 Material do Flange Flange Material - Indicates the material of the flange. R/W Enum 2048
ao_ring_material 4009 Material original Oring Material - Indicates the construction material of a seal that exists between the module and the flange. R/W Enum 2048
adrain_vent_material 400A Material de ventilação do material Drain Vent Material - Indicates the material of the vents. R/W Enum 2048
aremote_seal_isolator_material 400B Material do isolador de selo remoto Remote Seal Isolator Material - Defines the material of the remote seal wetted diaphragm. R Enum 2048
amodule_isolator_material 400C Material isolador Module Isolator Material - Indicates the material of the sensor diaphragm. R Enum 2048
aremote_seal_type 400D Tipo de selo remoto Remote Seal Type - Defines the type of Remote seal. R Enum 2048
remote_seal_fill_fluid 400E Fluído preenchido com selo remoto Remote Seal Fill Fluid - Defines the remote seal filling liquid. R Enum 2048
anumber_of_remote_seals 400F Número do selo remoto Number of Remote Seals - Physical quantity of the seals for the remote seal that is installed. R Enum 2048
amodule_fill_fluid 4010 Preencher fluído Sensor Fill Fluid - Fluid material in the module. R Enum 2048
module_range_code 4011 Faixa de módulo Module range - This is the range for the sensor. R Enum 2048
acalibration_type 4012 Tipo de calibração Calibration Type - Indicates if std. or specific on operating pressure or temperature or both. R Enum 2048
procedure_type 4013 Procedure Type Procedure Type - It is the type of any special procedure required for the manufacture of the sensor, e.g. degreasing. R Enum 2048
local_keys_mode 4014 Local keys mode Local keys mode control - Disables the Zero and Span Push buttons. R/W Enum 2048
lcd_display_mode 4015 LCD display mode LCD display mode control - Selects the variables shown on the integral display. R/W Enum 2048
root_transfer_mode 4016 Root transfer mode Root transfer function mode - selects behaviour of square root output at input values from 0 - 4 %. R/W Enum 2048
apressure_value 4018 Pressão Field Device Primary Variable Value - Represents the input pressure. R Float 10
sensor_limit_units 4019 Sensor Limits Units Sensor Limit Units and Minimum Span Units - Engineering units that the Upper Sensor Limit, Lower Sensor Limit and Minimum Span will be displayed with. R Enum 8
amodule_type 401A Tipo de modulo Module Type - Defines the Module type of the field instrument. R Enum 2048
special_transfer_function 401B Special Transfer Function Special Transfer Function. R Enum 2048
lower_sensor_trim_point 401C Sensor baixo de ponto de ajuste Lower Sensor Trim Point - Applied process pressure at which the last lower sensor trim was performed. Pressure Sensor - Pressure hardware that measures pressure with respect to some reference pressure. R/W Float 2048
Name ID Label Description Type Access Class