Loading…
Name ID Label Description Type Access Class
display_2nd_line 4030 2nd line The process display for the 2nd line of the secondary. R/W Enum 2048
display_1st_line_multipl 4031 1st line multipl. The multiplex mode for the 1st line of the secondary. R/W Enum 2048
display_2nd_line_multipl 4032 2nd line multipl. The multiplex mode for the 2nd line of the secondary. R/W Enum 2048
operating_mode 4024 Operating mode Select STANDARD for continuous flow. Select FAST for piston pump operation and short time fills >3 s. 2 RANGE, AUTOMATIC sets function P1 P2 to RANGE. MIN- and MAX-Alarm always refer to range 1. 2 RANGE, EXTERN sets function X1 to EXT.RANGE. MIN- and MAX-Alarm always refer to the selected range. R/W Enum 2048
flowdirection 4025 Flowdirection flowdirection is selected in forward and reverse flow or forward flow only. R/W Enum 2048
flowindication 4026 Flowindication Reverses the flow direction designations. R/W Enum 2048
module 403C Module Selects the hardware module placed in the secondary R/W Enum 2048
range_velocity 4050 Range velocity Selects range velocity in m/s or ft/s. R/W Enum 2048
range_05_DN 4051 range<.05rangeDN Ability to select upper range value less than .05 rangeDN. R/W Enum 2048
range_fixed 4052 range DN Range DN set to FIXED allows the selection of METER SIZE from the implemented list. Range DN set to programable allows direct input to Range DN, metersize is not selectable. R/W Enum 2048
Name ID Label Description Type Access Class