Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
filter | 4022 | Filter | Select Filter on, if the output signal is noisy or for pulsating flows (pulsation free output). | R/W | Enum | 2048 |
prog_unit | 4035 | Prog unit | select with density if the programable unit is massflowand without density if it is volumeflow. | R/W | Enum | 2048 |
function_p1p2 | 4029 | Function P1/P2 | on operating mode 2 RANGE,AUTOMATIC terminal P1/P2 signals the flow range. On the other operating modes all other functions are available. | R/W | Enum | 2048 |
function_x1 | 4023 | Function X1 | on operating mode 2 RANGE,EXTERN terminal X1 selets the flow range. If the contact is open range 1 is selected, if it is closed range 2 is selected. On the other operating modes the other functions are available. | R/W | Enum | 2048 |
Iout_at_alarm | 402D | Iout at Alarm | Select HIGH or LOW current output for error conditions. | R/W | Enum | 2048 |
detector_e_pipe | 4011 | Detector e.pipe | select ON if the detector e.pipe shall signals an empty pipe. | R/W | Enum | 2048 |
alarm_e_pipe | 4012 | Alarm empty pipe | If set to ON an empty pipe is signaled with the combined output signal. If set to off no contact output signal will be set. | R/W | Enum | 2048 |
Iout_at_e_pipe | 4013 | Iout at empty pipe | Select HIGH or LOW current output on empty pipe. | R/W | Enum | 2048 |
totalizer_function | 4016 | Totalizer funct. | If STANDARD is selected the count pulses for forward and reverse flow directions are integrated separately in two different counters. On DIFFER.TOTAL only a single difference totalizer is utilized for both flow directions. On forward flow the totalizer is incremented, on reverse flow he is decremented. | R/W | Enum | 2048 |
display_1st_line | 402F | 1st line | The process display for the 1st line of the secondary. | R/W | Enum | 2048 |
Name | ID | Label | Description | Type | Access | Class |