Name ID Label Description Type Access Class
konst_druck 404A const. Pressure Constant value for pressure. Used if pressure measurement is not available or in off mode. R/W Float 2048
zeit_freq_halten 405B Frequ. hold time Time to hold last measured value if there are bad quality signals with drop outs. R/W Float 2048
abgl_Iout_4mA 4053 Adjust Iout 4 mA Adjust 4 mA current output value. R/W Float 2048
abgl_Iout_20mA 4054 Adjust Iout 20 mA Adjust 20 mA current output value. R/W Float 2048
abgl_Iin_4mA 4055 Adjust Iin 4 mA Adjust 4 mA current input value for pressure transmitter. R/W Float 2048
abgl_Iin_20mA 4056 Adjust Iin 20 mA Adjust 20 mA current input value for pressure transmitter. R/W Float 2048
k_mittel 405D Mean k-factor K-factor linearisation in off mode. Use the mean k-factor of calibration. R/W Float 2048
k1 405F k1 k-factor k1 for k-factor linearisation. Please refer to calibration protocol! R/W Float 2048
k2 4061 k2 k-factor k2 for k-factor linearisation. Please refer to calibration protocol! R/W Float 2048
k3 4063 k3 k-factor k3 for k-factor linearisation. Please refer to calibration protocol! R/W Float 2048