Loading…
Loading…
Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
tv_upper_limit | 412A | TV USL | Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. | R | Float | 8 |
tv_lower_limit | 412B | TV LSL | Upper Sensor Limit Value- Defines the maximum usable value for the Upper Range Value. | R | Float | 8 |
tv_minimum_span | 412C | TV Min span | Minimum Span Value- Defines the smallest allowable difference between the Upper Range Value and the Lower Range Value. | R | Float | 8 |
tv_damping_value | 412D | TV Damp | Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. | R | Float | 8 |
tv_serial_number | 412E | TV Snsr s/n | Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. | R | Unsigned | 2048 |
tv_UpdatePeriod | 412F | TV Update Period | R | Time | 0 | |
qv_value | 4130 | QV (Deviation) | Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention | R | Float | 10 |
qv_units | 4131 | QV units | Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit | R/W | Enum | 8 |
qv_classifi_code | 4132 | QV Classifi Code | Classification- Indicates the unit code expansion table used for a particular dynamic or device variable. | R | Enum | 0 |
qv_PDQ | 4133 | QV PDQ | Process Data Quality- Indicates the overall process data quality of the device variable value. | R | Enum | 10 |
Name | ID | Label | Description | Type | Access | Class |