Loading…
Name ID Label Description Type Access Class
sv_damping_value 4120 SV Damp Damping Value- Damping time constant applied to both the Digital Value representation and Analog Value of the respective Analog Output. The units of this variable are always in seconds. R Float 8
sv_serial_number 4121 SV Snsr s/n Sensor Serial Number- Uniquely identifies the sensor from which the Digital Value representation or Field Device variable is primarily derived. R Unsigned 2048
sv_UpdatePeriod 4122 SV Update Period R Time 0
tv_value 4123 TV (Temperature) Field Device Variable Value- Digital value that changes magnitude with respect to time with no user intervention R Float 10
tv_units 4124 TV Unit Field Device Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit R/W Enum 8
tv_classifi_code 4125 TV Classifi Code Classification- Indicates the unit code expansion table used for a particular dynamic or device variable. R Enum 0
tv_PDQ 4126 TV PDQ Process Data Quality- Indicates the overall process data quality of the device variable value. R Enum 10
tv_LS 4127 TV LS Limit Status- Indicates whether the device variable is limited (i.e., not responding to a process). R Enum 10
tv_FStatus 4128 TV Fmly stat Device Family Status- Indicates additional status information on a device variable for a specific device family. R BitEnum 10
tv_DFamily 4129 TV Dev fmly Device Family- Indicates the family (i.e, pressure, temperature, flow, etc.) that the device variable belongs. R Enum 2048
Name ID Label Description Type Access Class