Variable Name: Word_420

Variable ID: 0x40C1

Variable Label: Status Word 420

Variable Help: Reads Modbus register no. 420

Variable Handling: R

Variable Class: Dynamic | Correction

Variable Size: 2

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Primary mA Output Saturated
0x0002 Secondary mA Output Saturated
0x0004 Primary mA Output Fixed
0x0008 Secondary mA Output Fixed
0x0010 Density Outside Limits
0x0020 Drive Over-Range / Partially Full Tube
0x0040 PIC/Daughterboard Communication Failure
0x0080 External Input Failure
0x0100 Core EEPROM Checksum Error (Config, Powerdown, Program)
0x0200 Core RAM Error
0x0400 Sensor Not Responding (No Tube Interrupt)
0x0800 Temperature Sensor Out-of-Range
0x1000 Input Over-Range
0x2000 Frequency Output Saturated
0x4000 Transmitter Not Characterized (Flocal or Sensor Type)
0x8000 Real-Time Interrupt Failure

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
215 40B2 read_additional_diags

Variable Write Command List

None