Loading…
Loading…
Variable Name: install_factor_B
Variable ID: 0x4039
Variable Label: B
Variable Help: Install Factors - A, B, C; Install Factors are utilized in situations when the measured flow rate may need to be adjusted due to variations in flow profile or other installation factors. The Install Factors are the coefficients of the polynomial equation: Corrected Flow = A + Bx + Cx^2 where x is the flow measured by the TA2 in PV units.
Variable Handling: R/W
Variable Class: Device
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|.6f
Variable Edit Format: |en|.6f
Variable Unit:
Variable Max:
Variable Min:
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
153 | 4127 | read_install_factors_A_B_C |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
154 | 4128 | write_install_factors_A_B_C |