Loading…

Variable Name: DielectricRange

Variable ID: 0x4686

Variable Label: Dielectric Range

Variable Help: A broad gain setting. The parameter selects the range bounding the dielectric constant of the medium being measured. Choose range 'Below 1.7' for very light hydrocarbons such as butane and propane. For stillwell applications only. Choose range '1.7 - 3' for typical hydrocarbons and oils. Choose range '3 - 10' for amines. Choose range 'Above 10' for water-based applications. If uncertain, chose lower range setting for additional signal strength. (Some ranges may not be selectable depending on the antenna model.)

Variable Handling: R/W

Variable Class:

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 Below 1.7
0x0001 1.7 - 3.0
0x0002 3 - 10
0x0003 Above 10

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
135 48A6 read_DielectricRange

Variable Write Command List

Command Number Command ID Command Name
136 48A7 set_DielectricRange