Variable Name: SZFault

Variable ID: 0x403C

Variable Label: SZ Fault State

Variable Help: Safety Zone Fault State defines how the Analog Output will respond when the transmitter detects a level within the Safety Zone. Enable fault to ensure safe, reliable high-level readings in critical applications. If Latch 3.6 or Latch 22 is chosen, the unit will remain in alarm until it is manually cleared with the SZ Alarm Reset.

Variable Handling: R/W

Variable Class: Diagnostic | AnalogOut

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 None
0x0001 3.6mA
0x0002 22mA
0x0003 Latch 3.6
0x0004 Latch 22

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
207 40C2 read_SZFault

Variable Write Command List

Command Number Command ID Command Name
208 40C3 set_SZFault