Loading…

Variable Name: DielRange

Variable ID: 0x468E

Variable Label: Dielectric Range

Variable Help: A broad gain setting. The parameter selects the range bounding the dielectric constant of the medium being measured. For interface measurement mode, it selects the range bounding the dielectric constant of the lower liquid medium. Choose range 'Below 1.7' for very light hydrocarbons such as butane and propane. Choose range '1.7 - 3' for typical hydrocarbons and oils. Choose range '3 - 10' for amines. Choose range 'Above 10' for water-based applications. If uncertain, chose lower range setting for additional signal strength. (Some ranges may not be selectable depending on the probe model.)

Variable Handling: R/W

Variable Class:

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 Below 1.7
0x0001 1.7 - 3.0
0x0002 3 - 10
0x0003 Above 10

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
135 49AE read_DielRange

Variable Write Command List

Command Number Command ID Command Name
136 49AF set_DielRange