Loading…

Variable Name: device_status

Variable ID: 0x0097

Variable Label: Device status

Variable Help: Field Device Status- Data indicating that the Field Device has detected condition(s) in the Field Device relating to: its hardware / the validity of a variable / its operating status / an internal process

Variable Handling: R

Variable Class: Dynamic | Diagnostic | HART

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Process applied to the primary variable is outside the operating limits of the field device
0x0002 Process applied to the non-primary variable is outside the operating limits of the field device
0x0004 PV Analog Channel Saturated
0x0008 PV Analog Channel Fixed
0x0010 Field device has more status available
0x0020 A reset or self test of the field device has occurred, or power has been removed and reapplied
0x0040 A modification has been made to the configuration of the field device
0x0080 Field device has malfunctioned due to a hardware error or failure

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
523 1303 read_condensed_status_map
523 1303 read_condensed_status_map
523 1303 read_condensed_status_map
523 1303 read_condensed_status_map
204 4847 rcmBe32EventsStatus
130 487D rcmGdcState
140 487E rcmFlGlobalLower
141 4881 rcmFlGlobalUpper
144 4883 rcmColFlUniGlobalLower
148 4885 rcmCol2FlUniGlobalLower
149 4888 rcmCol2FlUniGlobalUpper
152 488B rcmUiGlobalLower
152 488B rcmUiGlobalLower
153 488D rcmUiGlobalUpper
156 4891 rcmUi16GlobalLower
156 4891 rcmUi16GlobalLower
160 4894 rcmCol2FlGlobalLower
161 4897 rcmCol2FlGlobalUpper
164 489B rcmStGlobalLower
164 489B rcmStGlobalLower
165 489E rcmStGlobalUpper
168 48A2 rcmDatTimGlobal
172 48A6 rcmCNumber
172 48A6 rcmCNumber
174 48A7 rcmElectronicSerialNumber
176 48A9 rcmColFlUniLocal
185 4957 rcmAccessLevelHart
192 4958 rcmCustodyLock
187 495A rcmColFlUniGdcFbOutput
188 495B rcmIdIdentnumberAndHardwareCode
189 495C rcmIdSoftwareVersionNumber
190 495D rcmIdProductionAndCalibrationDate
191 495E rcmIdBoardSerialNumber
210 4960 rcmUi32GlobalLower
0 00AE read_unique_identifier
1 00AF read_pv
2 00B0 read_pv_current_and_percent_range
3 00B1 read_dynamic_variables_and_pv_current
9 00DD read_device_variables_and_status
7 00DB read_loop_configuration
8 00DC read_dynamic_variable_classification
14 00B6 read_pv_sensor_info
15 00B7 read_pv_output_info
12 00B4 read_message
13 00B5 read_tag_descriptor_date
16 00B8 read_final_assembly_number
20 00DE read_long_tag
48 1043 read_additional_device_status
48 1043 read_additional_device_status
33 1054 read_device_variables
50 1052 read_device_variable_assignments
54 1057 read_device_variable_information
60 0FE5 read_analog_output_and_percent_of_range
62 1062 read_analog_channels
63 104B read_analog_channel_information
70 1022 read_analog_channel_endpoint_values
90 115E read_real_time_clock

Variable Write Command List