Loading…

Variable Name: device_specific_status_1

Variable ID: 0x102C

Variable Label: Status group warning 1

Variable Help: Status group warning 1

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0080 PV between 1.1 x FS and 1.5 x FS
0x0040 Electronic temperature out of limits (-45°...+125°)
0x0020 No output linearization
0x0010 No temperature compensation
0x0008 No pre linearization
0x0004 PV out of upper limit (1.5 x full scale value)
0x0002 Electronic temperature out of range (-40°...+85°)
0x0001 Internal communication failed

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None