Variable Name: mreFunc
Variable ID: 0x40B6
Variable Label: Mult. reflections
Variable Help: Multiple reflections frequently occur in storage tanks with an extremely calm surface, typically when the BM 70 A/P /700/702 is mounted on a dome or centred on a "dished cover", or the vessel is fitted with a flat or only very slightly convex head. Such multiple reflections tend to simulate lower-than-actual levels. In such cases, activate the multiple reflections identifier, and see also references in Sect. 8.8 of 'Operating Manual'. · NO (default) Measurement without identification of multiple reflections. · YES Measurement with identification of multiple reflections. A better solution, however, is to change the mounting location of the BM 70 A/P /700/702 so as to prevent the occurrence of multiple reflections in the first place, or at least to diminish the strength of the multiple reflections (see Sect. 6.1.2 of 'Operating Manual'). To ensure satisfactory identification of the first measured value after the BM 70 A/P /700/702 has been switched on, always carry out measurements with the "empty spectrum" when the "multiple reflection identifier" is activated, see Sect. 8.6.12 of 'Operating Manual'. (Fct. 3.5.5)
Variable Handling: R/W
Variable Class: Correction
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | No |
0x0001 | Yes |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
132 | 40AE | read_user_data_enums |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
133 | 40C6 | write_user_data_enums |