Variable Name: ftbFunc
Variable ID: 0x40BB
Variable Label: Function FTB
Variable Help: The tank bottom tracing system (see Sect. 8.6.15 of 'Operating Manual') is activated under this function: · OFF The FTB is deactivated; the device operates in the standard mode. · PARTIAL When this option has been selected, the tank bottom tracing system is activated only in the vicinity of the tank bottom, i.e. at low levels (max. 20% of the tank height but a minimum of 0.6 m). Given higher levels, the normal measuring method is used (reflection from the product surface). This method is sufficient for most applications with poorly reflecting products. The parameter "Epsilon R" must be as exact as possible. · FULL (for BM 70 A only) When this option is selected, the tank bottom tracing system is activated over the total measuring range (tank height). For tank heights above 16 m, however, measuring accuracy is additionally affected. The parameter "Epsilon R" must be as exact as possible. (Fct. 3.5.7)
Variable Handling: R/W
Variable Class: Device
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | Off |
0x0001 | Partial |
0x0002 | Full |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
132 | 40AE | read_user_data_enums |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
133 | 40C6 | write_user_data_enums |