Loading…
Loading…
Variable Name: range
Variable ID: 0x402A
Variable Label: Range
Variable Help: The RANGE setting is determined by the PROBE LENGTH. RANGE 1 is set for probes shorter than 100 feet. RANGE 2 is set for probes from 100 to 200 feet. Changing the RANGE setting will erase all factory trim settings and calibration values. Changing the RANGE setting is not recommended unless a commitment to changing the physical probe length has been made.
Variable Handling: R/W
Variable Class: Device
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | 1 Short Range |
0x0001 | 2 Long Range |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
151 | 4075 | read_basic_device_setup |
160 | 407B | read_extended_setup |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
152 | 4076 | write_probetype_mounting_config_flags |
162 | 407D | write_gain_and_config_flags |