Variable Name: pressure_damping_value
Variable ID: 0x1398
Variable Label: PV Damping [sec]
Variable Help: This allows filtering to be applied to the digital and analog representation of the Primary Variable. The damping value represents one time constant (i.e. response to step input reaches 63% of final steady-state value after time specified has elapsed) as per the HART specification. This damping is applied to both the digital Primary Variable as well as the analog Loop Current.
Variable Handling: R/W
Variable Class: Correction
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|5.2f
Variable Edit Format: |en|5.2f
Variable Unit:
Variable Max:
Variable Min:
Variable Read Command List
None
Variable Write Command List
None