Variable Name: device_status

Variable ID: 0x0097

Variable Label: Device status

Variable Help: Field Device Status- Data indicating that the Field Device has detected condition(s) in the Field Device relating to: its hardware / the validity of a variable / its operating status / an internal process

Variable Handling: R

Variable Class: Dynamic | Diagnostic | HART

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Process applied to the primary variable is outside the operating limits of the field device
0x0002 Process applied to the non-primary variable is outside the operating limits of the field device
0x0004 PV Analog Channel Saturated
0x0008 PV Analog Channel Fixed
0x0010 Field device has more status available
0x0020 A reset or self test of the field device has occurred, or power has been removed and reapplied
0x0040 A modification has been made to the configuration of the field device
0x0080 Field device has malfunctioned due to a hardware error or failure

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
128 4029 ReadLocationStringVariable
130 402D ReadVersion
136 402F ReadSerialNumber
137 403F ReadOneLowFloat
143 404C ReadOneLowUnsignedInteger
153 404F ReadProcessEvents
147 4053 ReadOneLowString
149 4066 ReadClassifiedStatus
180 40BA ReadLocationStringVariableSC200
182 40BF ReadVersionsAndSNSC200
183 40C2 ReadDiagnosticsSC200
186 40EE ReadOutput_1_SC200
187 40EF ReadOutput_2_SC200
189 4131 ReadOneLowFloatSC200
191 4182 ReadLowUnsignedIntegerSC200
192 4183 ReadRelayA_1_SC200
193 4184 ReadRelayA_2_SC200
194 4185 ReadRelayB_1_SC200
195 4186 ReadRelayB_2_SC200
196 4187 ReadRelayC_1_SC200
197 4188 ReadRelayC_2_SC200
198 4189 ReadRelayD_1_SC200
199 418A ReadRelayD_2_SC200
201 41A1 ReadHighFloatSC200_1
202 41A2 ReadHighFloatSC200_2
204 41C8 ReadHighUnsignedIntegerSC200
205 41CC ReadOneLowUnsignedLongIntegerSC200
206 41DA ReadHighUnsignedLongIntegerSC200
207 41DD ReadLowStringVariableSC200
220 4250 ReadAnalogChannelsHART
176 427E ReadHARTOutputLoopA
177 427F ReadHARTOutputLoopsB
178 4280 ReadHARTOutputLoopC
170 4285 ReadDeviceMappingVariableHARTModule
171 4286 ReadLocationStringVariableHARTModule
173 4288 ReadVersionsAndSNHARTModule
208 428A ReadDate
210 428D ReadTime
0 00AE read_unique_identifier
1 00AF read_pv
2 00B0 read_pv_current_and_percent_range
3 00B1 read_dynamic_variables_and_pv_current
9 00DD read_device_variables_and_status
7 00DB read_loop_configuration
8 00DC read_dynamic_variable_classification
14 00B6 read_pv_sensor_info
15 00B7 read_pv_output_info
12 00B4 read_message
13 00B5 read_tag_descriptor_date
16 00B8 read_final_assembly_number
20 00DE read_long_tag
48 1043 read_additional_device_status
48 1043 read_additional_device_status
50 1052 read_device_variable_assignments

Variable Write Command List

None