Loading…

Variable Name: device_status

Variable ID: 0x0097

Variable Label: Device status

Variable Help: Field Device Status- Data indicating that the Field Device has detected condition(s) in the Field Device relating to: its hardware / the validity of a variable / its operating status / an internal process

Variable Handling: R

Variable Class: Dynamic | Diagnostic | HART

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Process applied to the primary variable is outside the operating limits of the field device
0x0002 Process applied to the non-primary variable is outside the operating limits of the field device
0x0004 PV Analog Channel Saturated
0x0008 PV Analog Channel Fixed
0x0010 Field device has more status available
0x0020 A reset or self test of the field device has occurred, or power has been removed and reapplied
0x0040 A modification has been made to the configuration of the field device
0x0080 Field device has malfunctioned due to a hardware error or failure

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
128 4015 ReadLocationStringVariable
130 4019 ReadVersion
136 401B ReadSerialNumber
137 4021 ReadOneLowFloat
143 402C ReadOneLowUnsignedInteger
153 402F ReadProcessEvents
147 4035 ReadOneLowString
149 4048 ReadClassifiedStatus
180 4078 ReadLocationStringVariableSC200
182 407D ReadVersionsAndSNSC200
183 4080 ReadDiagnosticsSC200
186 40AC ReadOutput_1_SC200
187 40AD ReadOutput_2_SC200
189 40EF ReadOneLowFloatSC200
191 4140 ReadLowUnsignedIntegerSC200
192 4141 ReadRelayA_1_SC200
193 4142 ReadRelayA_2_SC200
194 4143 ReadRelayB_1_SC200
195 4144 ReadRelayB_2_SC200
196 4145 ReadRelayC_1_SC200
197 4146 ReadRelayC_2_SC200
198 4147 ReadRelayD_1_SC200
199 4148 ReadRelayD_2_SC200
201 415F ReadHighFloatSC200_1
202 4160 ReadHighFloatSC200_2
204 4186 ReadHighUnsignedIntegerSC200
205 418A ReadOneLowUnsignedLongIntegerSC200
206 4198 ReadHighUnsignedLongIntegerSC200
207 419B ReadLowStringVariableSC200
220 4204 ReadAnalogChannelsHART
176 4232 ReadHARTOutputLoopA
177 4233 ReadHARTOutputLoopsB
178 4234 ReadHARTOutputLoopC
170 4239 ReadDeviceMappingVariableHARTModule
171 423A ReadLocationStringVariableHARTModule
173 423C ReadVersionsAndSNHARTModule
208 423E ReadDate
210 4241 ReadTime
0 00AE read_unique_identifier
1 00AF read_pv
2 00B0 read_pv_current_and_percent_range
3 00B1 read_dynamic_variables_and_pv_current
9 00DD read_device_variables_and_status
7 00DB read_loop_configuration
8 00DC read_dynamic_variable_classification
14 00B6 read_pv_sensor_info
15 00B7 read_pv_output_info
12 00B4 read_message
13 00B5 read_tag_descriptor_date
16 00B8 read_final_assembly_number
20 00DE read_long_tag
48 1043 read_additional_device_status
48 1043 read_additional_device_status
50 1052 read_device_variable_assignments

Variable Write Command List

None