Variable Name: error_status
Variable ID: 0x402F
Variable Label: Error Status
Variable Help: Prioritized Unit Faults
Variable Handling: R
Variable Class: Dynamic | Correction
Variable Size: 2
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | |
0x0001 | CLOSE TO LOW IR |
0x0002 | DIRTY LENS |
0x0004 | LOW IR |
0x0008 | HIGH IR |
0x0010 | WIRE SHORT |
0x0020 | LOW LINE VOLTS |
0x0040 | FAILED TO CALIBRATE |
0x0080 | FAILED TO ZERO |
0x0100 | GAS CHECK TIME OUT |
0x0200 | OVER TEMPERATURE |
0x0400 | XMTR FAULT |
0x0800 | HEATER FAILURE |
0x1000 | SET UP MENU |
0x2000 | MISC FAULT |
0x4000 | EXCESS NEG. DRIFT |
0x8000 | NVM CHECKSUM FAULT |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
163 | 40A1 | get_fast_info |
Variable Write Command List
None