Loading…
Loading…
Variable Name: device_specific_status_19
Variable ID: 0x1036
Variable Label: Status Group 19
Variable Help: Field Device Specific Status Byte 20- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0004 | F170 Pressure cell connection defective |
0x0008 | S171 Ambient temperature too low |
0x0010 | S172 Ambient temperature too high |
0x0020 | S173 Sensor range exceeded |
0x0040 | F174 Pressure cell electronics defective |
0x0080 | M175 Pressure cell deactivated |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None