Loading…
Loading…
Variable Name: device_specific_status_1
Variable ID: 0x102C
Variable Label: Status Group 1
Variable Help: Field Device Specific Status Byte 2- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0002 | C485 Measured variable simulation Remedy information Deactivate simulation |
0x0004 | F022 Temperature sensor defective Remedy information 1. Check or replace sensor electronic module (ISEM) 2. If available: Check connection cable between sensor and transmitter 3. Replace sensor |
0x0008 | 046 Sensor limit exceeded Remedy information 1. Inspect sensor 2. Check process condition |
0x0010 | F910 Tubes not oscillating Remedy information 1. Check electronic 2. Inspect sensor |
0x0020 | C453 Flow override Remedy information Deactivate flow override |
0x0080 | 842 Process limit Remedy information Low flow cut off active! 1. Check low flow cut off configuration |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None