Loading…
Loading…
Variable Name: SPV_SimulateDiagCode_1
Variable ID: 0x442D
Variable Label: Diag. event sim.
Variable Help: Select a diagnostic event to simulate this event
Variable Handling: R/W
Variable Class: Device
Variable Size: 4
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x80EC | Off |
0xBE0000 | 022 Temperature sensor defective |
0xC80000 | 046 Sensor limit exceeded |
0xDB0000 | 062 Sensor connection faulty |
0x2B30000 | 063 Exciter current faulty |
0xE70000 | 082 Data storage |
0xA00000 | 083 Memory content |
0xCC0000 | 140 Sensor signal asymmetrical |
0x1C70000 | 144 Measuring error too high |
0x14B0000 | 201 Device failure |
0x670000 | 242 Software incompatible |
0x6B0000 | 252 Modules incompatible |
0x1490000 | 262 Sensor electronic connection faulty |
0x780000 | 270 Main electronic failure |
0x7D0000 | 271 Main electronic failure |
0x790000 | 272 Main electronic failure |
0x980000 | 273 Main electronic failure |
0x7A0000 | 275 I/O module 1 defective |
0x7A0001 | 275 I/O module 2 defective |
0x7A0002 | 275 I/O module 3 defective |
0x7B0000 | 276 I/O module 1 faulty |
0x7B0001 | 276 I/O module 2 faulty |
0x7B0002 | 276 I/O module 3 faulty |
0xE10000 | 283 Memory content |
0x1EE0000 | 302 Device verification active |
0x26C0000 | 303 I/O 1 configuration changed |
0x26C0001 | 303 I/O 2 configuration changed |
0x26C0002 | 303 I/O 3 configuration changed |
0xE20000 | 311 Electronic failure |
0x2C90000 | 330 Flash file invalid |
0x2CA0000 | 331 Firmware update failed |
0x2C70000 | 332 Writing in HistoROM backup failed |
0x950000 | 361 I/O module 1 faulty |
0x950001 | 361 I/O module 2 faulty |
0x950002 | 361 I/O module 3 faulty |
0xA10000 | 372 Sensor electronic (ISEM) faulty |
0x2D10000 | 373 Sensor electronic (ISEM) faulty |
0xCE0000 | 374 Sensor electronic (ISEM) faulty |
0x1070000 | 375 I/O-1 communication failed |
0x1070001 | 375 I/O-2 communication failed |
0x1070002 | 375 I/O-3 communication failed |
0x16D0000 | 382 Data storage |
0x16E0000 | 383 Memory content |
0x2880000 | 387 HistoROM backup failed |
0x8B0000 | 410 Data transfer |
0x2040000 | 412 Processing download |
0x40000 | 431 Trim 1 |
0x600000 | 437 Configuration incompatible |
0x6A0000 | 438 Dataset |
0x990000 | 441 Current output 1 |
0x940000 | 453 Flow override |
0x900000 | 484 Failure mode simulation |
0x930000 | 485 Measured variable simulation |
0xE0000 | 491 Current output 1 simulation |
0x2840000 | 502 CT activation/deactivation failed |
0x2760000 | 520 I/O 1 hardware configuration invalid |
0x2760001 | 520 I/O 2 hardware configuration invalid |
0x2760002 | 520 I/O 3 hardware configuration invalid |
0x3870000 | 528 Concentration settings faulty |
0x3890000 | 529 Concentration settings faulty |
0x14A0000 | 537 Configuration |
0x2070000 | 540 Custody transfer mode failed |
0x2820000 | 599 Custody transfer logbook full |
0xAD0000 | 803 Current loop |
0xC00000 | 830 Sensor temperature too high |
0xC20000 | 831 Sensor temperature too low |
0xC30000 | 832 Electronic temperature too high |
0xC10000 | 833 Electronic temperature too low |
0xC50000 | 834 Process temperature too high |
0xC60000 | 835 Process temperature too low |
0x910000 | 842 Process limit |
0x920000 | 862 Partly filled pipe |
0x310000 | 882 Input signal |
0x500000 | 910 Tubes not oscillating |
0xC40000 | 912 Medium inhomogeneous |
0xCD0000 | 913 Medium unsuitable |
0x3800000 | 941 API temperature out of specification |
0x33B0000 | 942 API density out of specification |
0x37F0000 | 943 API pressure out of specification |
0x1C60000 | 944 Monitoring failed |
0x1680000 | 948 Oscillation damping too high |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
173 | 4618 | read_index_parameter_173 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
178 | 4616 | write_index_parameter_178 |