Loading…

Variable Name: device_specific_status_0

Variable ID: 0x102B

Variable Label: Status Group 0

Variable Help: Field Device Specific Status Byte 1- Data that has Field Device status information relating to: failures / warnings / status of process's.

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0008 231 Date/time not available -> Remedy information: 1. Replace display module or its cable 2. Set date and time
0x0010 C431 Trim 1 required -> Remedy information: Carry out trim
0x0040 C491 Current output 1 simulation active -> Remedy information: Deactivate simulation

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None