Loading…
Loading…
Variable Name: HBM_BurstDeviceVariableCode4_1
Variable ID: 0x41B7
Variable Label: Burst variable 3
Variable Help: For HART command 9 and 33, assign a HART device variable or process variable to burst variable
Variable Handling: R/W
Variable Class: Device
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | Volume flow |
0x0001 | Mass flow |
0x0002 | Corrected volume flow |
0x0003 | Flow velocity |
0x0007 | Electronic temperature |
0x0009 | Totalizer 1 |
0x000A | Totalizer 2 |
0x000B | Totalizer 3 |
0x0008 | Density |
0x0006 | Temperature |
0x000F | HART input |
0x00F4 | Percent of range |
0x00F5 | Measured current |
0x00F6 | Primary variable (PV) |
0x00F7 | Secondary variable (SV) |
0x00F8 | Tertiary variable (TV) |
0x00F9 | Quaternary variable (QV) |
0x00FA | Not used |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
173 | 4585 | read_index_parameter_173 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
178 | 4583 | write_index_parameter_178 |