Loading…
Loading…
Variable Name: device_specific_status_2
Variable ID: 0x102D
Variable Label: Status Group 2
Variable Help: Field Device Specific Status Byte 3- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0008 | F242 Software incompatible Remedy information 1. Check software 2. Flash or change main electronics module |
0x0010 | F261 Electronic modules Remedy information 1. Restart device 2. Check electronic modules 3. Change I/O Modul or main electronics |
0x0020 | F262 Module connection Remedy information 1. Check module connections 2. Change electronic modules |
0x0040 | F437 Configuration incompatible Remedy information 1. Restart device 2. Contact service |
0x0080 | M438 Dataset Remedy information 1. Check data set file 2. Check device configuration 3. Up- and download new configuration |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None