Loading…

Variable Name: parActualError

Variable ID: 0x407B

Variable Label: present error

Variable Help:

Variable Handling: R

Variable Class: Dynamic

Variable Size: 2

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000
0x0065 A101 checksum error general reset & new calibr.required
0x0066 A102 checksum error general reset & new calibr.required
0x0067 W103 initialising - please wait
0x006A A106 downloading please wait
0x006E A110 checksum error general reset & new calibr.required
0x006F A111 electronics defect
0x0071 A113 electronics defect
0x0072 A114 electronics defect
0x0073 A115 fault power supply
0x0074 A116 download error repeat download
0x0079 A121 electronics defect
0x0098 A152 checksum error general reset & new calibr.required
0x0099 W153 initialising - please wait
0x009B A155 electronics defect
0x00A0 A160 checksum error general reset & new calibr.required
0x00A4 A164 electronics defect
0x00AB A171 electronics defect
0x00E7 A231 sensor 1 defect check connection
0x010E A270 custody switch undef check position
0x010F W271 reduced accuracy check installation
0x0110 A272 electronics defect amplifier
0x0111 W273 reduced accuracy
0x0112 W274 reduced accuracy check pipe diameter
0x0113 W275 electronics defect factory settings
0x01FF W511 no factory cali- bration ch1
0x0200 W512 recording of mapping please wait
0x0259 W601 linearisation ch1 curve not monotone
0x0263 W611 less than 2 linea- risation points for channel 1
0x026D W621 simulation ch. 1 on
0x0281 E641 no usable echo channel 1 check calibr.
0x0283 E643 no usable echo channel 1 check calibr.
0x028B E651 level in safety distance - risk of overspill
0x029F A671 linearisation ch1 not complete, not usable
0x02A9 W681 current ch1 out of range

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

None

Variable Write Command List