Variable Name: xmtr_specific_status_4
Variable ID: 0x0FB9
Variable Label: Status group 5
Variable Help: Field Device Specific Status Byte 5- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0001 | E115: electronics defect |
0x0002 | E164: electronics defect |
0x0004 | E155: electronics defect |
0x0008 | E110: checksum error general reset + new calibr.required |
0x0010 | E171: electronics defect |
0x0020 | E1121: Current output not calibrated |
0x0040 | W153: initialising - please wait |
0x0080 | W103: initialising - please wait |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 0FCE | read_additional_transmitter_status |
Variable Write Command List
None