Loading…

Variable Name: device_specific_status_0

Variable ID: 0x102B

Variable Label: Specific Status

Variable Help: Specific Statuses Related to Analog Output Current, Sensor and Measurement.

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Measured values are not ready/Diagnostics
0x0002 Sensor value is out of range
0x0004 Poor sensor signal
0x0008 Sensor temperature is out of range
0x0010 Flow cutoff
0x0020 Sensor communication error
0x0040 One or more parameters are corrupted
0x0080 Service maintenance

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None