Loading…
Loading…
Variable Name: xmtr_specific_status_0
Variable ID: 0x0FB5
Variable Label: Status group 1
Variable Help: Field Device Specific Status Byte 1- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0001 | Alarm Code 1, Checksum error- Field device EEPROM checksum verification failure. |
0x0002 | Alarm Code 2, DPM error- DPM reports status error. |
0x0004 | Alarm Code 4, DPM information error- DPM information packets have not been received (on power up). |
0x0008 | Alarm Code 8, DPM connection failure- Communications with DPM timed out. |
0x0010 | Alarm Code 16, DPM information mismatch- Packet does not match first received information. |
0x0020 | Alarm Code 32, DPM packet fault- Bad command, byte count, or CRC in DPM packet. |
0x0040 | Alarm Code 64, EEPROM write error- Failure to write to field device EEPROM. |
0x0080 | Fld dev stat1-8 |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 0FCE | read_additional_transmitter_status |
Variable Write Command List
None