Loading…

Variable Name: vMBErrE4SincePowerOn

Variable ID: 0x4301

Variable Label: Sensor Error E4 since Power On

Variable Help:

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 MB Fehler (group error)
0x0002 Checksum error in Dräger data
0x0004 Checksum error in user data, device init required
0x0008 Checksum error in Dräger data sensor
0x0010 Checksum error in user data sensor, device init required
0x0020 No sensor connected
0x0040 Wrong sensor version connected
0x0080 Wrong sensor type connected

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
205 42F9 read_error_history

Variable Write Command List

None