变量名称: x3302_middle_oi
变量ID: 0x404B
变量标签: Middle Oi
变量描述: Percent of calibrated Oi test signal returned by the middle sensor. Reading should be 90 to 110 percent. If reading is below 90 percent, first try clean the window if the reading is still below 90 go to (Device Setup), select (Calibration) and perform Oi Calibration. If over 110% recalibrate.
变量处理: 读
变量分类: 动态 | 诊断 | 设备
变量长度: 2
变量数据类型: 无符号
变量显示格式:
变量编辑格式:
变量单位:
变量最大值: N/A
变量最小值: N/A
变量读取命令列表
命令号 | 命令ID | 命令名称 |
---|---|---|
160 | 4065 | x3302_cmd_rd_middle_oi |
变量写入命令列表