Loading…

Variable Name: transfer_function

Variable ID: 0x1F84

Variable Label: Xfer fnctn

Variable Help: Transfer Function- Defines the transformation function that will be applied from the Field Device Variable to the Analog Output, and Percent Range. The Transfer Function does not affect the Digital Value representation.

Variable Handling: R

Variable Class: InputBlock

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 Linear
0x0001 Sq root
0x0002 Sq root to 3rd power
0x0003 Sq root to 5th power
0x0004 Spcl curve
0x0005 Square
0x00E6 discrete (switch)
0x00E7 Sq root+ spcl curve
0x00E8 Sq R to 3rd P+SCurv
0x00E9 Sq R to 5th P+SC
0x00FA Not used
0x00FB None
0x00FC Unknown
0x00FD Spcl

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
15 00B7 read_pv_output_info

Variable Write Command List

None