Loading…

Variable Name: EmRateDesired

Variable ID: 0x43F4

Variable Label: Desired Firing Rate

Variable Help: The desired emission rate or time between the firing of two transducers in sequence based on the firing order. The actual emission rate used will not be less than the meter's calculated minimum based on the meter's geometry (pipe diameter, distance between transducers, axial distance between transducers , the minimum speed of sound and the physical characteristics of the transducers themselves. The actual emission rate used may also be impacted by stacking. A value of (0) ensures the use of fastest available rate determined by the meter.

Variable Handling: R/W

Variable Class:

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|.6g

Variable Edit Format: |en|.6f

Variable Unit:

Variable Max:

Variable Min:

Variable Read Command List

Command Number Command ID Command Name
185 4402 ReadAcquisitionConfiguration

Variable Write Command List

Command Number Command ID Command Name
184 43F0 WriteAcquisitionConfiguration