Loading…

Variable Name: xmtr_specific_status_4

Variable ID: 0x0FB9

Variable Label: Status group 5

Variable Help: Field Device Specific Status Byte 5- Data that has Field Device status information relating to: failures / warnings / status of process's.

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0080 Alarm 16: A/D convertor defective
0x0040 Info: Channel 1 not ready
0x0020 Alarm 53: Outside of sensor 1 upper range
0x0010 Alarm 53: Outside of sensor 1 lower range
0x0008 Alarm 51: Sensor 1 short circuit
0x0004 Alarm 50: Sensor 1 open circuit
0x0002 Alarm 52: Sensor 1 corrosion
0x0001 Warning 206: Sensor 1 corrosion

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 0FCE read_additional_transmitter_status

Variable Write Command List

None