Variable Name: MeasurementPh_asymmetricPot
Variable ID: 0x4012
Variable Label: Asymmetric Pot
Variable Help: The Asymmetric Potential value is the electrical potential across the measuring and reference half-cell of the sensor at the isopotential point. Can be modified only for a Custom pH sensor Type. Valid only when Measurement Type = pH.
Variable Handling: R/W
Variable Class:
Variable Size: 4
Variable DataType: Float
Variable Display Format: |en|.2f
Variable Edit Format: |en|8.2f
Variable Unit:
Variable Max:
Variable Min:
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
213 | 4243 | ReadPVSensorConfig_Ph_213 |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
217 | 4247 | WriteIsopotAsym_Ph_217 |